CH26303 หลอด Tube วัดสารเคมีประเภท Organic Arsenic Compounds and Arsine ยี่ห้อ Draeger
Model : 75-2104
Model (old) : 62-2000-O04
Application Range
Standard Measuring Range: Qualitative 0.1 ppm Arsine and 3 mg org. arsenic/m3 are the minimum
detectable concentrations.
Number of Strokes n: 8
Time for Measurement: max. 3 min
Standard Deviation: ± 50 %
Color Change: yellow -> grey
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: < 50 mg H2O / L
Reaction Principle
a) AsR3 + Zn/HCl -> AsH3
b) AsH3 + Au/Hg-complex -> Au (colloidal)
Cross Sensitivity
Phosphine and arsine are indicated before the ampoule is broken, but with different sensitivities.
Additional Information Arsine is present if a grey ring appears in the indicating layer after performing 8 pump strokes. If there is no indication, the ampoule must be broken and the liquid transferred onto the indicating layer such that it is completely saturated. Then an additional eight pump strokes must be performed.